| Dimensions | 300 mm x 250 mm x 75 mm |
| Weight | 2 kg net |
| Gross Weight and Size (Packed) | Weight: 4 kg |
| Packed Size | 470 mm x 112 mm x 377 mm |
| Subjects | Automatic population of database from Spirotrac. Name, ID; Age, Height, Gender; Smoking Status, Body Mass Index and more depending on model. Storage of up to 10,000 subjects |
| Selectable Parameters (depending on model) | VC, IVC, IC, VT (TV), TLC, RV, IRV, ERV, FRC, FVC, FIVC, FEV1, FEV3, FEV6, FVC, FEV1/VC, FEV1/FVC, FEV3/VC, FEV3/FVC,FEV1/FEV6, FEF75, FEF50, FEF25, FEF25-75, FEF25-75/FVC, FIV1, PIF, FIV1/FIVC, FIF25, FIF50, FIF75, FEF50/FIF50, FET, MVVind, FEV1 Ratio, FEV0.5, PEF L/min, PEF L/s, FEF 0.2-1.2, FEF 75-85, FEF25, FEF50, FEF75, FMFT, FIF25, FIF50, FIF75, PIF, FEV0.75, Lung Age and more. |
| Environmental Data | Selectable: Temperature (built in sensor), Barometric pressure, altitude and humidity |
| Test Types | Single breath tests, flow/volume loops, multi-breath testing, tidal breathing and combined VC/FVC type test methods supported |
| Predicted Values Selectable (depending on model) | ERS/ECCS; NHANES, Polgar, Pereira, Berglund, Forsche, Gutierrez, Hedenström, Taiwan; Knudson, Crapo; Hsu; KNLW; Viljanen; SEPAR; Gulsvik; SBPT; Tamura; Ip; Quanjer; Wang; Dockery and more |
| Suggested Interpretation | User selectable |
| Storage Humidity | 10% to 95% |
| Storage Temperature | 0 °C to 50 °C |
| Recommended Operating Temperature Range | 17°C–37 °C; Design Limits: 10 °C–40 °C |
| Data Storage | Stores up to 10,000 subjects |
| PowerSAFE™ | Input 100–240 V AC 50-60 Hz, output 12 V DC |
| Battery Pack | Rechargeable NiMh, 7.2 V 1, 800 mAH |
| Flow Detection Principle | Fleisch type pneumotachograph |
| Volume Detection | Flow integration sampling @ 100 Hz |
| Resolution | 10 mL volume; 0.01 L/s flow |
| Accuracy when Operated in Operating Temperature Range Conditions | Flow ± from 10% to 5%, Max flow rate ±16L/s, Min flow rate ±0.02L/s |
| Volume Accuracy | Better than ±3% |
| Maximum Displayed Volume | 10 L |
| Linearity | ±3% |
| Maximum Test Duration | 90 seconds |
| Flow Impedance | <0.1 kPa/L/s @ 14 L/s Complies with ATS/ERS 2005 |
| Printer | Thermal |
| Performance Standards | ISO 26782:2009, ISO 23747:2007; ATS/ERS 2005 |
| Safety Standards | IEC 60601-1:2005 |
| Medical Devices Directive | 93/42/EEC1993 as amended |
| Designed & Manufactured Under | ISO 13485:2003, FDA 21CFR820 and CMDR |
| Communications | USB and CF card |
| Connectivity | USB 2 B |